Warning: To ensure correct nebulisation, maintain the nebuliser in a
vertical orientation (Figure 13 & Figure 14).
Use with a Nasal Interface
The Aerogen Solo can be used on/off ventilator with a nasal interface
when configured with a humidifier (Figure 7).
Aerogen Ultra
The Aerogen Ultra is an accessory specific to the Aerogen Solo nebuliser.
It facilitates intermittent and continuous nebulisation, with optional supply
of supplemental oxygen to paediatric and adult patients via mouthpiece.
The device can alternatively be used with a face mask, as supplied.
It is a single patient use device which is qualified for 20 intermittent use
treatments (at a rate of four 3 mL doses per day over 5 days) or 3 hours of
continuous use.
Optimal aerosol delivery is achieved with a valved mouthpiece or valved
aerosol face mask with low/no oxygen flow.
Inspect for device integrity and correct valve placement prior to use.
Mouthpiece
Face Mask
Aerogen Ultra
Aerogen Solo
Oxygen Tubing
Figure 15. Assembly of Aerogen Ultra
23