Use only with Aerogen Solo components, connectors and any accessories,
which are specified by Aerogen in this instruction manual.
Do not use beyond defined life (see page 33).
Do not use in the presence of flammable substances or flammable
anaesthetic mixtures combined with air, oxygen or nitrous oxide.
To avoid the risk of fire, do not use to aerosolise alcohol-based medications,
which can ignite in oxygen-enriched air and under high pressure.
Do not autoclave any component or accessory of the Aerogen Solo
System.
Do not modify this equipment without the authorisation of the manufacturer.
Do not use or store outside of specified environmental conditions.
To avoid damage to the nebuliser:
• Do not apply undue pressure to the domed aperture plate in the centre
of the nebuliser.
• Do not push out the Aerogen Vibronic
• Do not use a syringe with a needle to add medication.
• Do not attempt to clean the nebuliser.
Condensate can collect and occlude ventilator and/or patient circuits.
Always position ventilator and/or patient circuits so that fluid condensate
drains away from the patient.
8