Functional Test

Perform a functional test of the Aerogen Solo System prior to first use or
at any time to verify proper operation. This test is to be carried out prior to
inserting the nebuliser into a circuit or accessory.
1. Visually inspect each part of the system for cracks or damage and
replace if any defects are visible.
2. Pour 1-6 mL of normal saline (0.9%) into the nebuliser.
3. Connect the nebuliser to the controller using the controller cable.
Connect the AC/DC adapter to the controller and plug the AC/DC
adapter into an AC power source.
4. Press and release the blue On/Off power button and verify that the
green 30 Min. indicator light illuminates and that aerosol is visible.
5. Disconnect the nebuliser from the controller. Verify that the amber
Error Indicator lights. Reconnect the nebuliser to the controller.
6. Disconnect the AC/DC adapter from the controller and verify that
nebulisation continues and that the battery status indicator turns off.
7. Power off the controller. Reconnect the AC/DC adapter to the controller.
Press and hold the button for at least 3 seconds. Verify that the green
Continuous indicator light illuminates and that aerosol is visible.
8. Turn the system off and verify that the 30 Min. and Continuous
indicators are off.
Aerogen® Solo System Instruction Manual