Table 2. Aerogen Pro-X Controller Troubleshooting (Continued)
If this happens:
Medication is left in
the nebuliser after
nebulisation cycle.
Flashing amber light.
Note: The rechargeable battery in the Aerogen Pro-X Controller should
only be replaced by Aerogen authorised personnel: contact your Aerogen
sales representative.
32
It could mean:
Nebuliser was not
turned on or connected
to power.
Rechargeable battery is
depleted.
A 30 Minute cycle
was selected when
connected to the
continuous feed system.
It may be time to replace
the nebuliser.
It may mean that it
is time to replace
controller.
Aerogen®
Try this:
Ensure that nebuliser is
connected to power and
turned on.
Recharge battery (see
Recharging the Battery).
Run a continuous cycle.
See Warranty and Life
of Product. Refer to the
Aerogen Solo parts list by
visiting www.aerogen.com.
Contact your local Aerogen
sales representative.