Aerogen Solo Aerosol Flow Rate Calculation - Aerogen Solo Manuel D'utilisation Du Système

Masquer les pouces Voir aussi pour Solo:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Aerogen Solo Aerosol Flow Rate
Calculation (Optional)
Flow rates may vary between individual Aerogen Solo nebulisers. The
minimum flow rate for all Aerogen Solo nebulisers is 0.2 mL per minute.
In order to calculate the flow rate of an individual Aerogen Solo nebuliser,
follow these steps:
1. Transfer 0.5 mL of normal saline (0.9%) or intended drug into the
Aerogen Solo medication cup.
2. Turn on the nebuliser.
3. Using a stop-watch, measure the length of time it takes from the start
of nebulisation until all the saline/drug has been nebulised.
4. Calculate the flow rate using the following equations:
Flow rate in mL/min =
Flow rate in mL/hr =
30
)
Volume of normal saline or drug
Nebulisation time in seconds
)
)
Volume of normal saline or drug
Nebulisation time in seconds
Aerogen®
)
x 60
)
)
x 60
x 60

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Pro-x

Table des Matières