Table 11.
Guidance and manufacturer's declaration - electromagnetic immunity for
the Aerogen Solo nebuliser system that is not life supporting
This Aerogen Solo nebuliser system is intended for use in the electromagnetic
environment specified below. The customer or the user of the Aerogen Solo nebuliser
system should assure that it is used in such an environment.
IEC 60601 Test
Immunity Test
Level
Conducted RF
3 Vrms outside
IEC 61000-4-6
industrial,
EN 61000-4-6:
scientific and
2014
medical (ISM)
and amateur
radio bands. 6
Vrms in ISM and
amateur radio
bands
150 kHz to 80
MHz
Note:
Aerogen® Solo System Instruction Manual
Compliance
Electromagnetic
Level
Environment - Guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the Aerogen Solo nebuliser
system, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
10 Vrms
Recommended Separation
150 kHz to 80
Distance
MHz
d = [1.17] √P
45