Cleaning the Aerogen Pro-X Controller
Cleaning of controller and controller cable, AC/DC adapter and mounting
brackets:
1. Wipe clean with an alcohol based disinfectant wipe or a quaternary
ammonium compound based disinfectant wipe.
2. Check for exposed wiring, damaged connectors, or other defects and
replace controller if any are visible.
3. Visually inspect for damage and replace the controller if any damage
is observed.
Warnings
• Do not immerse or autoclave the Aerogen Pro-X Controller, cable or
AC/DC adapter.
• Do not place the Aerogen Pro-X Controller in an incubator during use.
• Do not use abrasive or sharp tools.
• Do not spray liquid directly onto the controller.
• Do not wrap the nebuliser cable tightly around any of the system
components.
• Do not use in the presence of devices generating high electromagnetic
fields such as magnetic resonance imaging (MRI) equipment.
• The Aerogen Pro-X Controller contains a nickel metal hydride (NiMH)
rechargeable battery, which should be disposed of in accordance with
local governing regulations at the end of its useful life.
• Follow local laws and recycling plans regarding disposal or recycling of
components, batteries and packaging.