1. Insert the Aerogen Solo nebuliser firmly into the Aerogen Ultra in
orientation shown in Figure 15.
2. If supplemental oxygen is required, firmly attach oxygen tubing to the
Aerogen Ultra.
Note: Oxygen flow rate should be set between 1-6 LPM.
3. If a face mask is required, remove mouthpiece and attach face mask
to the Aerogen Ultra.
Note: When using an open face mask, a minimum oxygen flow of 1 LPM
is required.
4. Add medication to nebuliser.
5. Connect cable to the Aerogen Solo and power on controller.
6. Introduce the Aerogen Ultra to patient and observe aerosol flow to
ensure correct operation.
7. Remove excess rainout from the Aerogen Ultra periodically (hourly with
continuous nebulisation).
8. To ensure optimum performance of the Aerogen Ultra, remove any
residue by rinsing through with sterile water, shake off excess and allow
to air dry.
Warnings
• Do not use with a closed face mask.
• When using with an open face mask, always use supplemental oxygen
flow of 1-6 LPM.
• Performance of the Aerogen Ultra may vary depending upon the type
of drug and the Aerogen Ultra configuration used.
• Do not exceed recommended oxygen flow for system.
• Ensure oxygen connection port or tubing is not occluded.
• Do not use the Aerogen Ultra without a mouthpiece or face mask.
• Visually check the Aerogen Ultra post-rinsing to ensure that valves
have not become dislodged.
• Do not cover the Aerogen Ultra valves during use.
• Do not use the Aerogen Ultra in conjunction with the Aerogen Pro.
• Do not autoclave any component of the kit.
• Ensure tubing is safely orientated to prevent strangulation hazard.