Installation For Use With A Ventilator - Aerogen Solo Manuel D'utilisation Du Système

Masquer les pouces Voir aussi pour Solo:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Installation for use with a Ventilator

T-Pieces - Connection to a Breathing Circuit
1. For 22mm adult breathing circuits connect the nebuliser with adult
T-piece into the inspiratory limb of the breathing circuit before the
patient Y (Figure 5).
Figure 5. Connecting the Aerogen Solo to an adult breathing circuit
For 15mm paediatric breathing circuits connect the nebuliser with
the paediatric T-piece into the inspiratory limb of the breathing circuit
before the patient Y, as shown for the adult T-piece in Figure 5.
The Aerogen Solo can connect to 10mm neonate breathing circuits
with the 15mm paediatric T-piece and the neonate adapters. This can
be positioned approximately 30cm (12 in.) back from the patient Y
(Figure 6).
Figure 6. Connecting to a neonate breathing circuit
Aerogen® Solo System Instruction Manual
15