10 MAINTENANCE

10.7 Maintaining the Base Adjustment

The base adjustment should not require any attention other than:
1.
Check the squareness of the legs when in the closed position.
2.
Place a square on the inside of the legs and base to determine the
90° alignment.
3.
Adjust the linkage rods until 90° alignment is achieved.

10.8 Replacing Casters/Forks

Replacing Rear Casters

1.
Place the lift on its side.
2.
Remove the bolt and locknut that secure the existing rear caster to
the fork.
The bushing will be loose and may fall out of the caster.
Existing bushing will be reused. Examine and replace if
worn.
3.
Install the new/existing bushing into the new rear caster.
4.
Line up the mounting holes in the new rear caster and the fork.
5.
Install the bolt through the fork and new rear caster and tighten
securely with the locknut.
Manual/Electric Portable Patient Lift
Linkage Rod
Locknut
Rear Caster
Bushing
48
Square
Linkage Rods
90°
Base
Fork
Bolt
Part No. 1078987