To carry out a single-gas calibration:
1. Mount the calibration adapter to the gas detector. Make
sure that both guide pins are in the correct grooves. Avoid
putting the pump adapter at the wrong angle.
2. Connect the test gas cylinder to the calibration adapter.
3. Connect a second hose (max. length: 2 m) to the second
connector on the calibration adapter to direct the test gas
to an exhaust or outside.
4. Switch on the gas detector.
> Login.
5. Select
6. Enter and confirm the password.
7. Select Maintenance > Single gas cal..
 A dialog for selecting the measuring channel to be
calibrated appears.
8. Select the measuring channel.
 A dialog for selecting the calibration appears.
9. Select either zero-point calibration or span calibration.
● For a span calibration: Enter and confirm the
calibration concentration.
10. Open the valves on the test gas cylinder.
11. Select
to start the single-gas calibration or select
abort the calibration.
 The measuring channel appears, and the measured
value flashes.
As soon as the stability check detects a stable
measured value, a calibration is carried out
automatically.
12. If necessary, press
this case, a calibration takes place immediately.
 The new measured value is displayed for control
purposes.
The result is displayed as follows:
single-gas calibration successful.
single-gas calibration failed.
13. The single-gas calibration is complete when measuring
channel successfully passed or failed the single-gas
calibration.
● Select
and then confirm the dialog to discard the
result.
● Select
to confirm the result.
14. Close the valves on the test gas cylinder.
15. Remove the calibration adapter.
16. After completing the measurement, make sure that the
seal on the calibration adapter is clean and that there are
no chips of metal on the seal.
Plug the protective cap onto the calibration adapter to
protect the seal from damage and being deformed.
If there was an error during the single-gas calibration:
– Repeat the single-gas calibration.
– Inspect the sealing contours and surfaces on the
calibration adapter as well as the front cradle of the
housing to make sure they are free of damage. Inspect the
thread port for the calibration adapter.
– If necessary, replace the sensor.
For mixed gas and cross calibration, see the technical
manual.
Instructions for use
|
X-am
to overrule the stability control. In
®
3500
5.6

Charging the battery

WARNING
Risk of explosion!
To reduce the risk of ignition of a flammable or explosive
atmosphere, observe the following:
► Do not open the gas detector in explosion-hazard areas.
► Only use the LBT 02** (lithium-ion) battery type.
► Do not charge or exchange the battery in explosion-
hazard areas.
► Only use the battery charger specified by Dräger. The use
of a different charger nullifies the explosion protection
certification of the gas detector.
For the replacement of the battery, see the technical
manual.
The battery is a part of the lower part of the casing. The
battery can be charged with or without the gas detector.
1. Place the gas detector or only the lower part of the casing
with the battery in the charging cradle.
 The gas detector switches off automatically (factory
to
setting). The green LED on the power pack flashes.
The typical charging time after a work shift of 8 – 10 h is
approx. 4 h.
The typical charging time for an empty battery is approx. 10 h
If the battery is deep-discharged, it may be necessary to
leave the device in the charging cradle for up to 16 h to
charge it.
If the specified temperature range (5 to 35 °C) is not reached
or is exceeded, charging stops automatically. This extends the
charging time. After a return to the temperature range,
charging continues automatically.
WARNING
No measurement!
If there is an external power supply voltage drop > 1 s in the
charging cradle while the gas detector is in operation, the gas
detector switches off.
► Ensure uninterrupted power supply (does not apply if the
optional setting, that the gas detector does not
automatically turn off in the charging cradle, is selected). If
this cannot be ensured, regularly check that the gas
detector is turned on (e.g. based on the optical and
acoustic operating signals).
Designation and descrip-
tion
Inductive charging cradle for
charging 1 gas detector
Adapter for the plug-in power
supply unit
Plug-in power supply unit for
charging 1 gas detector
Plug-in power supply unit for
charging 5 gas detectors
|
en
Maintenance
Order no.
83 25 825
83 25 736
83 16 997
83 16 994
39