WARNING
Incorrect device feature/settings!
Incorrect device features/settings may result in danger to life
and/or risk of explosion.
► Before every use, check whether the display elements, the
alarm features and information are displayed correctly. If
one of the items listed above does not feature correctly or
is incorrect, do not use the gas detector and have it
inspected.
The following features are active during the sensor warm-up
phase:
– The measured values flash
– The yellow LED is illuminated
– A warning notice is displayed
 The gas detector is ready to measure when the measured
values no longer flash and the yellow LED is no longer
illuminated. The warning notice may continue to be
displayed if warnings are pending. For further information
see: "Opening information", page 33.
No alarms are issued during the warm-up phase!
4.4.3 Switching off the gas detector
1. Hold down ▲ and ▼ simultaneously until the displayed
countdown expires.
 The visual, acoustic and vibration alarms activate
briefly.
 The gas detector is switched off.
Or
1. Select
in measuring mode and confirm the dialog.
2. Select and confirm Switch off.
The gas detector may only be switched off without a prior
sign-on if the Switch-off allowed feature is activated using
the Dräger CC-Vision PC software. Factory setting: activated
The gas detector switches off automatically when it is
placed in the charging cradle (factory setting). As an option,
this setting can be deactivated using the Dräger CC-Vision
PC software.
If the gas detector remains off for longer than 21 days and
is not charged, deep sleep mode is activated. In deep sleep
mode, the gas detector can no longer be switched on using
the Dräger CC-Vision PC software or the Dräger X-dock. In
this case, the gas detector must be switched on manually. The
sensors then go through their warm-up phase.
4.5

Log user in or out

The gas detector has four configurable user levels. The user
levels can be set using the Dräger CC-Vision PC software.
User level 0 means that the user is not signed on. User levels
1 to 3 require a password for sign-on.
The following passwords are created by default:
User level 1:
Instructions for use
|
X-am
0001
®
3500
User level 2:
User level 3:
Dräger recommends changing the predefined passwords
after initial commissioning.
Default setting:
Feature
Bump test
Fresh air calibration
1)
Settings menu
1)
Maintenance menu
1) Maintenance and settings menu in user level 0 are not part of the
technical suitability test.
To log in a user:
1. Select
in measuring mode and confirm the dialog.
2. Select and confirm Login.
3. Enter the four-digit user level password and confirm each
digit.
To sign out a user:
1. Select
in measuring mode and confirm the dialog.
2. Select 'Sign out' and confirm the dialog.
4.6

Preparations for operation

WARNING
Serious damage to health
An incorrect calibration can lead to incorrect measured
values, which may result in serious damage to health.
► Before performing safety measurements, check the
calibration using a bump test, adjust as necessary, and
check all alarm elements. If national regulations exist, the
bump test must be performed in accordance with these
regulations.
► If the unit is permanently in the charging cradle while
being operated, then the bump test must be carried out
again after 4 weeks at the latest and also if the location of
the device is changed.
WARNING
Risk of explosion
Only applicable if using the Dräger CSE Connect app:
The use of an unsuitable smartphone or accessories may
lead to the ignition of flammable or explosive atmospheres.
► The smartphone on which CSE Connect is installed must
be suitable and approved for use in explosion-hazard
areas.
► A list of compatible smartphones is available from Dräger.
► Only use accessories that are suitable for use in areas
subject to explosion hazards.
Operation
0002
0003
User level
0
-
-
-
-
-
-
-
|
en
-
-
-
-
31