SAFETY STANDARDS

Follow all the instructions given in this manual.
The autoclave must be used by personnel specifically trained in the sterilisation process, within medical clinics,
chiropody studios and beauty salons, under the responsibility of the director of health and the sterilisation process
supervisor.
Only personnel who have been properly trained on the contents of this instruction manual may use the autoclave. The
medical facility in which the autoclave is installed is responsible for the training on the product operation and main-
tenance. It must keep training records and check that personnel have effectively understood the training.
The list of organisations that are authorised to carry out maintenance and repairs is available in the SUPPORT/LIST
OF SUPPORT CENTRES section on the www.faro.it website. The user is responsible for requesting servicing from
authorised centres only.
There are no procedures that the operator can perform to check that the product is working correctly.

1.1 Intended users

medical staff (dentists, chiropodists)
dental hygienists
employees in beauty salons
USER PROFESSIONAL QUALIFICATIONS
Degree in medicine
Dental hygienist
MINIMUM SKILLS
Those required by the professional qualification
Course on use based on this manual for unqualified users
LANGUAGE COMPREHENSION
Those acquired for the professional qualification
EXPERIENCE
That required to carry out the profession
POSSIBLE USER HANDICAPS
Full use of an upper limb is required
Eyesight compatible with the profession

1.2 Intended use

It is a small steam autoclave for sterilising tools, rotary instruments, reusable medical devices and porous materials
(e.g. fabrics) using saturated steam at a temperature of 121°C or 134°C.

1.3 General safety warnings

The product must be used for its intended use in accordance with the provisions of this manual.
Any use that does not comply with the instructions in the instruction manual could compromise the safety level of
the device.
Never modify the equipment without written authorisation from FARO SpA.
The user is responsible for installation, training users, maintenance and any legal requirements related to the device,
including the periodic checks described in this manual.
Faro shall not be liable for harm to things and/or people caused by failure to use the product correctly as described in
this manual. Failure to comply with the provisions of this manual will void any liability of the manufacturer and the
product warranty.
The device discharges air from a small hole at its bottom rear. Do not obstruct this vent (17 fig. 2).
Do not place the autoclave on an unstable table or surface.
Fire or overheating hazard warnings
Only put products that are resistant to steam sterilisation inside, in accordance with the manufacturer's
provisions.
Do not place the device on a flammable surface.
The device is not suitable for use in oxygen-rich or potentially explosive atmospheres.
Do not block the air vents on the equipment casing during installation or operation.
Comply with the minimum distance requirements in detail if the autoclave is recessed.