VOLTCRAFT VC731 Notice D'emploi page 75

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Peak Hold, Peak Max/Min
VC731 and VC732
Measuring type
ACV
ACA
Accuracy defined for sine curve, ACV >5 Vrms
ACA >=5 Arms (50 – 400 Hz)
Frequency
VC731 and VC732
Range
20.00 – 99.99 Hz
20.0 – 999.9 Hz
0.020 – 9.999 kHz
Sensitivity:
Harmonic distortion
VC731 and VC732
Measuring type
ACA/ACV
Measurement for harmonic distortion
VC731 and VC732
Order
H01 – H12
H13 – H25
If the measured value ACV <10 Vrms or ACA <10 Arms, the LC display shows "rdy".
At a frequency outside of 45 - 65 Hz, the LC display indicates "outF".
Range
140.0 V
1400 V
140.0 A
850 A
Accuracy
±0.7 % +5 dgt
10 – 100 Vrms for AC 100 V
10 – 100 Arms for AC 100 A (>400 Hz)
100 – 1000 Vrms for AC 1000 V
100 – 600 Arms for AC 600 A (>400 Hz)
Range
99.9 %
Range
99.9 %
Accuracy
±4.3 % +22 dgt
±4.3 % +22 dgt
Accuracy
±4.3 % +15 dgt
Accuracy
±7.1 % +15 dgt
±14.3 % +15 dgt
75