Maintenance And Cleaning; Fuse Replacement; Disposal - VOLTCRAFT TC-150 Manuel D'utilisation

Temperature calibrator
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

MAINTENANCE AND CLEANING

Apart from an occasional cleaning or exchanging the fuse, this device is maintenance-free. Use a
clean, lint-free, antistatic and dry cloth to clean the device. Do not use any abrasive or chemical agents
or detergents containing solvents.
Let the device cool down fully before cleaning. There is a danger of burns when
touching the heating and cooling surface.

Fuse Replacement

The heating element and the control electronics in the device have separate fuses.
Fuse (12) for heating element, fi ne-wire fuse 6.3 x 30 mm fast-blowing F1.5 A/250 V
Fuse (13) for control electronics, fi ne-wire fuse 6.3 x 30 mm fast-blowing F500mA/250 V
Proceed as follows to replace the fuse:
- Switch off the device and unplug the mains cable.
- Turn the respective fuse holder from the casing.
- Replace the defective fuse with a new fuse of the same type and rated current.
- Screw on the fuse holder carefully.
- The device may be taken into operation again.

DISPOSAL

Old electronic devices are raw materials and should not be disposed of in the household
waste. At the end of its service life, dispose of the device at the community collection
points according
to the applicable statutory regulations. It is prohibited to dispose of the device in the
household waste.
20