General Instructions

1. Place the Inogen One® G3 in a
well ventilated location.
Air intake and exhaust must have clear
access. Locate the Inogen One® G3 in such a
way that any auditory alarms may be heard.
WARNING Avoid use of the Inogen One® G3 in presence of pollutants, smoke
or fumes. Do not use the Inogen One® G3 in presence of flammable
!
C
anesthetics, cleaning agents or other chemical vapors.
CAUTION
Do not obstruct air intake or exhaust when operating the device.
Blockage of air circulation or proximity to a heat source may lead to
internal heat buildup and shutdown or damage to the concentrator.
CAUTION
The Inogen One® G3 Concentrator is designed for continuous use. For
optimal sieve bed (columns) life, the product should be used frequently.
2. Ensure both particle filters are in place.
CAUTION
Do not operate the Inogen One® G3 without both particle filters in place.
Particles drawn into the system may damage the equipment.
Operating Instructions
5 Setting Model
Exhaust
Operating Instructions Chapter 3
4 Setting Model
Exhaust
Intake
Intake
11