Inogen One® G3 Column Change Procedure (continued)
11. Insert column (metal tube)
into the Inogen One® G3
concentrator. Do not leave the
column exposed; it should be
inserted into the Inogen One
G3 as soon as the dust caps
have been removed.
12. Push the column (metal tube) into the device
such that the column (metal tube) is fully seated
into the Inogen One® G3 concentrator. The spring
loaded latch button should fully return to the
closed position.
13. Repeat steps 10-12 to install the other column
(metal tube).
14. Connect the AC power supply cord to the Inogen
One® G3 and plug the power supply AC cord into an
electrical outlet. Do Not Power on the Inogen
One® G3 concentrator.
15. Press and hold the light button for 10 seconds and
screen will display "sieve reset" message. Release
button once message is displayed on screen.
16. P ress the bell button once and screen will display "sieve reset success".
17. Press the power button to turn on the Inogen One® G3, and use normally.
WARNING Do not use any columns other than those specified in this user manual.
The use of non-specified columns may create a safety hazard and/or
!
C
impair equipment performance and will void your warranty.
Closed and locked
Cleaning, Care and Maintenance Chapter 6
35