The Inogen One® G3 monitors various parameters during operation and utilizes an
intelligent alarm system to indicate a malfunction of the concentrator. Mathematical
algorithms and time delays are used to reduce the probability of false alarms while still
ensuring proper notification of an alarm condition.
If multiple alarm conditions are detected, the highest priority alarm will be displayed.
The following notification messages are accompanied by a single, short beep.
Message Display & Text
Please Wait
Shutting Down
HH:MM
Vx.x:Serial Number
Low Priority Alerts
The following low priority alert messages are accompanied by a double beep and a
solid red or yellow light.
Message Display & Text
Battery Low
Attach Plug
Replace Columns
or
O2 Service Soon
24
Condition/Action/Explanation
On/Off button has been pressed for two seconds.
Concentrator is performing system shut down.
Audible Alarm button has been pressed for five seconds.
Condition/Action/Explanation
Battery power is low, with less than 10 minutes remaining.
Attach external power supply or power down and insert
a fully charged battery.
Column maintenance is required within 30 days. Contact
your equipment provider to arrange for service.