Siemens SITRANS FUE1010 IP40 NEMA 1 Guide Rapide page 27

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Note
Minimum Ltn 18mm (0.75 in)
Installation Procedure (See figure for reference)
1. On a flat surface, attach the Spacer Bar to a Mounting Frame so that the Reference Hole
2. Slide the second Mounting Frame onto the other end of the Spacer Bar and align the
3. Wrap a Mounting Strap around the pipe. Make sure to position it so there is easy access
4. At the mounting location, place the Mounting Frame/Spacer Bar Assembly on the pipe so
5. Engage the end of the Mounting Strap with the Mounting Strap Adjusting Screw.
6. Slide strap under the spring clip of one of the Mounting Frames.
7. Tighten the Mounting Strap Screw enough to take up all of the slack, but not enough to
8. Rotate the assembly on the pipe to the final conditioned location, ensuring that it is straight
9. Tighten the mounting straps to seat the assembly firmly on the pipe. Do not over tighten.
FUE1010 IP40 (NEMA 1) Energy Portable
Compact Operating Instructions, 5/2009, A5E02384186-03
on the Spacer Bar fits over the metal post on the platform of the frame. Tighten the clamping
screw.
Number Index Hole with the metal post on the platform. Then tighten the clamping screw.
Ensure that the angled sides of both frames face away from each other.
to the Mounting Strap Adjusting Screw.
that it rests on the top of the pipe.
prevent rotation of the assembly.
along the pipe axis. (Refer to the sensor orientation diagram.)
Repeat procedure for the other Mounting Frame .
Commissioning
3.3 Sensor Installation
25

Publicité

Table des Matières
loading

Table des Matières