Problems
The irrigation pump goes on and
off erratically causing abnormal
pressure fluctuations.
There is too much water in the
pressure chamber.
There is insufficient pressure in
the system.
109076-KIM Rev B.indd 29
Potential Causes
Water may have entered the
pressure sensing line.
An air leak in the line may have
developed.
An irrigation problem exists.
29
Possible Solutions
Change the One Day Set
Irrigation tube if water has entered
the pressure sensing line. If
problem persists, contact DePuy
Mitek Customer Service.
Check that the luer-lock on the
irrigation and pressure sensing
tube are closed. If necessary,
change the irrigation tubing and
start again.
Check the stop-cock on the
arthroscope, and the clamps
under the saline bags and under
the pressure chamber.