Chapter 4. Pump System Functions
4.1. Run/Stop
Press the Run/Stop pad
to start all pump system
functions.
Note: numbers in ( ), refer to diagram on page 12.
109076-KIM Rev B.indd 24
4.2. Preset Pressure
The pressure is displayed
in brackets (eg. <50>) on
the LCD (5) for 2 seconds
when one of the pressure
pads is pressed, then the
actual pressure is displayed
(pressure settings are in
levels).
Level 50, equals 5 feet
of gravity pressure or the
equivalent of 112.5 mmHg
(±10%).
The pump will maintain the
pre-set level. When operating,
use the "+" and "–" pads (9)
to increase and decrease the
level. The factory preset level
is 50.
24
4.3. Filling the Chamber
Press the Fill Chamber pad
to fill the pressure chamber
initially, and throughout the
surgery as needed. The
chamber should never be over
1/3 full.