Pull on the tube and close the
tension rocker arm against the
tube.
Green LED blinks. Press Run/
Stop. The green LED blinks
initially, then remains on.
Warning:
It is mandatory that the pressure chamber remains vertical during the procedure and that the
pump be positioned at the same level as the patient.
Over filling the chamber may result in the filter (shown in 3.2.A) becoming wet prior to the pressure
transducer and therefore interfere with accurate pressure measurement. Should over filling of the chamber
occur, see Chapter 6 for Troubleshooting.
109076-KIM Rev B.indd 18
Position the pressure chamber
in the holder. Connect the
transducer to the pump by
firmly screwing it in clockwise.
Press the Fill chamber
pad several times until the
chamber is 1/3 full. The
green LED blinks until pad is
pressed.
18
Close the cover.
9/26/12 4:02 PM