Chapter 5. Maintenance: Cleaning &
Decontamination
Cleaning the FmS SOLO:
At the end of the operating day, switch the FMS
SOLO off, unplug the power cord and discard all
tubing. Wipe off the instrument's sides, front panel
and covers with a soft cloth that has been slightly
moistened with a neutral-pH detergent. Wipe again
using distilled water to rinse.
W arning:
DO NOT IMMERSE. DO NOT AUTOCLAVE. Do
not use flammable agents when cleaning and
disinfecting the FMS SOLO.
109076-KIM Rev B.indd 27
Foot Pedal: Cleaning and
Decontamination
After each operation, clean the foot pedal and the
power cord with a neutral-pH detergent. The foot
pedal board can be immersed, except the electrical
connections, which must remain dry.
W arning:
DO NOT IMMERSE ELECTRICAL CONNECTOR
OF FOOT PEDAL. DO NOT AUTOCLAVE. Do
not use flammable agents when cleaning and
disinfecting the Foot pedal.
Do not twist the cable when storing.
Do not unplug instrument by pulling on cable.