Chapter 2. Operating Precautions, Checks & Warnings; Storage And Handling - DePuy Mitek FMS SOLO Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 114
Chapter 2. Operating Precautions,
Checks and Warnings
Warnings:
Factory Settings: Factory settings are based on
experience; they should only be used as guidelines.
The surgeon is responsible for settings depending
on the surgical procedure performed.
Electrical safety testing should be performed by a
biomedical engineer or other qualified person.
Risk of electrical shock: Do not remove the cover.
Refer servicing to your DePuy Mitek Representative
or DePuy Mitek Customer Service in the USA at +1
(800) 382-4682, Option 2.
Do not connect the device to a power source that is
not properly earthed (grounded).
Disconnect the device from the main power source
when cleaning, servicing, or inspecting.
Inspect all equipment and cables periodically for
wear. Contact your DePuy Mitek Representative or
DePuy Mitek Customer Service in the USA at +1
(800) 382-4682, Option 2 if damage is noted.
Avoid fluid contact with the FMS SOLO, and its
electrical connectors.
Do not use flammable agents when cleaning and
disinfecting the FMS SOLO.
To avoid risk of fire, replace fuses with same type
and rating.
109076-KIM Rev B.indd 15

2.1. Storage and Handling:

Storage temperature: Do not expose the FMS
SOLO to temperatures above 50°C (122°F) or
below –10°C (14°F).
Operating Conditions: Do not operate the pump
in conditions above 40°C (104°F) or below 10°C
(50°F).
Damage: Do not use a FMS SOLO that has been
dropped or shows signs of damage.
Disposal: Do not incinerate.
Contact your DePuy Mitek Representative or
DePuy Mitek Customer Service in the USA at
+1 (800) 382-4682, Option 2 to arrange for safe
disposal.