E. Set-up for Next Operation.
The One Day Set Irrigation Tube can be left in place at the end of the previous operation with the Sterile
Buffer Zone One-Way Valve/Intermediary Tubing with One-Way Check Valve attached.
After checking that the luer-
lock is tightened, the sterile
nurse takes the irrigation tube
and passes the blue end (D)
to a non-sterile person.
Warning:
Never attach one Sterile
buffer Zone One-Way
valve/Intermediary
Tubing with One-Way
Check valve to another.
109076-KIM Rev B.indd 22
The non-sterile person
removes the old sterile buffer
zone by disconnecting the
luer-locks of the One Day Set
Irrigation Tube (C) and the
Intermediary irrigation tube...
22
... and immediately connects
the blue luer-lock (D) of the
new irrigation tube to the blue
luer-lock (C) of the One Day
Set Irrigation Tube.