b. Installing the Intermediary Irrigation Tube Ref. 281103, 281142, or 281103*
From the One Day
Set Irrigation Tube
Ref. 284504 or 284503
This tube connects the irrigation tube to the sheath in a sterile manner.
* R ef. 281103 does not include the Sterile Buffer Zone One-Way Valve/Intermediary Tubing with One-Way
Check Valve. Ref. 281103 is intended for single use of the One Day Set Irrigation tube.
After checking that the luer-
lock is properly tightened, the
sterile nurse takes the irrigation
tube and passes the blue end
(D) to a non-sterile person.
The pump must be in the Run mode (10), with the light on.
Prime the irrigation tube by opening the clamp under the
pressure chamber. When primed, close the clamp or the stop-
cock on the arthroscope (in the sterile zone).
109076-KIM Rev B.indd 19
Sterile Buffer Zone One-Way
Valve/Intermediary Tubing with
(D)
One-Way Check Valve
Blue
White
(D)
The non-sterile person
removes the blue caps from
the irrigation tube (D) and
the One Day Set Irrigation
Tube (C)...
Blue
(C)
19
Red
To the
arthroscope
Sheath
(C)
(D)
... and immediately
connects the One Day Set
Irrigation Tube (C) luer-lock, to
the irrigation tube (D) luer-lock.
9/26/12 4:02 PM