Fault detected: The motor stops suddenly
Possible cause
•Micro motor/Cable damaged
•Wrong voltage.
•Footswitch damaged.
•PCB damaged.
Fault detected: The Motor will not turn off
Possible cause
•Keyboard and/or footswitch
damaged
•Broken connections inside.
•PCB damaged
•Wrong voltage.
X – ELECTROMAGNETIC COMPATIBILITY
Important:
The power cord, micro-motor cord, and footswitch cord must be kept apart.
The I-Endo Dual requires special precautions to be taken with regard to electromagnetic compatibility. It
must be installed and prepared for use as described in chapter 4.
Certain types of mobile telecommunication devices such as mobile telephones are likely to interfere
with the I-Endo Dual. The recommended separation distances in this paragraph must therefore be
complied with.
The I-Endo Dual must not be used near to or on top of another device. If this cannot be avoided, its
operation under the conditions of use must be checked beforehand.
The use of accessories other than those specified or sold by the manufacturer as replacement parts may
have the consequence of increasing the emissions or decreasing the immunity of the I-Endo Dual.
Solution
Check to see that the micro motor cable is plugged into the
connector on the front right of the central unit, disconnect and
reconnect. Start the motor by pressing the [Motor ON/OF] button
on the keyboard. If the motor turn on, immediately discontinue use
of the device and contact Satelec after-sales-service.
Verify that the voltage matches the power supply. If it does, try to
switch the central unit off and then on again, or connect the device
to another electrical outlet.
Attempt to start the micro motor by using the [Motor ON/OFF
button] on the central unit keyboard. If unit does not turn on,
contact Satelec after-sales-service.
If you cannot find the cause of the malfunction, immediately
discontinue use of the device and contact Satelec after-sales-
service.
Solution
Try to turn off the motor by pressing the [Motor ON/OFF] button
on the keyboard or by using the footswitch.
If the motor will not turn off, immediately discontinue use of the
device and contact Satelec after-sales-service.
Immediately discontinue use of the device and contact Satelec
after-sales-service.
Verify that the voltage matches the power supply. If it does, try to
switch the central unit off and then on again, or connect the device
to another electrical outlet.