Satelec Acteon I-Endo Dual Manuel D'utilisation page 20

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 28
Fault detected: The display is not working properly
Possible cause
•Contra angle damaged.
•Wrong voltage.
•Possible interference.
•Broken connections inside.
•Display damaged.
•PCB damaged.
Fault detected: The keyboard is not working properly
Possible cause
•Keyboard damaged.
•Broken connections inside.
Fault detected: The sound signals are not working properly
Possible cause
•Broken connections inside
Fault detected: The motor doesn't work
Possible cause
• Micro motor cable not connected
properly
•Contra angle defective or not
properly calibrated.
•Footswitch damaged.
•Wrong voltage.
•PCB damaged.
•Motor damaged.
•Connector pins and/or cable
damaged.
Solution
Check the integrity of the contra angle.
Verify that the voltage matches the power supply. If it does, try to
switch the central unit off and then on again, or connect the device
to another electrical outlet.
Check to see if the central unit has been positioned too close to
another electrical device which can cause interference. In this
case, move the interfering electrical device further away from the
central unit.
If you cannot find the cause of the malfunction, immediately
discontinue use of the device and contact Satelec after-sales-
service.
Solution
Contact Satelec after-sales-service.
Solution
Immediately discontinue use of the device and contact Satelec
after-sales-service.
Solution
Check to see that the micro motor cable is plugged into the
connector on the front right of the central unit, disconnect and
reconnect.
Check the integrity of the contra angle. Remove the contra angle
and perform calibration.
Attempt to start the micro motor by using the [Motor ON/OFF
button] on the central unit keyboard. If unit does not turn on,
contact Satelec after-sales-service.
Verify that the voltage matches the power supply. If it does, try to
switch the central unit off and then on again, or connect the device
to another electrical outlet.
Immediately discontinue use of the device and contact Satelec
after-sales-service.
20

Publicité

Table des Matières