Immunity test
Electrostatic
6kV contact
discharge
8kV air
(ESD) EN61000-4-2
Burst/Fast
2kV for power
transient
supply lines
EN 61000-4-4
Surge
1kV differential
EN 61000-4-5
mode
2kV common
mode
Voltage dips,
5%Ut
short
(95% dip in U
interruptions and
0,5 cycle
voltage variation
on power supply
40% Ut
input lines.
(60% dip in U
cycle
70% U
(30% dip in U
25 cycle
5% U
EN 61100-4-11
(95% dip in U
5 sec
Power frequency
(50/60 Hz)
magnetic field
3A/m
EN 61100-4-8
I-Endo Dual is intended for use in the electromagnetic environment specified below. The I-Endo Dual
should only be used in such an environment.
Test level
Compliance level
EN 60601-1-2
6kV
8kV
2kV for power
supply lines
1kV differential
mode
2kV common
mode
5%Ut
) for
(95% dip in U
t
for 0,5 cycle
40% Ut
) for 5
(60% dip in U
v
5 cycle
70% U
t
) per
(30% dip in U
t
25 cycle
5% U
t
) for
(95% dip in U
t
3A/m
Electromagnetic environment -
Floor should be wood, concrete or
ceramic tile. If floor are covered with
synthetic material, the relative humidity
should be at least 30%
Quality of the Main power source should
be that of a typical commercial or
hospital environment
Quality of the Main power source should
be that of a typical commercial or
hospital environment
Quality of the Main power source should
be that of a typical commercial or
)
t
hospital environment If the user of the I-
Endo Dual requires continued operation
during power main interruptions, it is
recommended than the I-Endo Dual be
) for
v
powered from an uninterruptible power
supply or battery.
t
) per
t
t
)
t
for 5 sec
Power frequency magnetic fields should
be at a typical commercial or hospital
establishment level.
guidance
23