Immunity Test
Conducted RF
EN 61000-4-6
Radiated RF
IEC 61000-4-3
Field strengths from fixed RF transmitter, as determined by an electromagnetic site survey
compliance level in each frequency range
following symbol:

10.3 RECOMMENDED SEPARATION DISTANCES

Recommended separation distance between portable and mobile RF communications equipment and the I-Endo
The I-Endo Dual is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the I-Endo Dual can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the I-Endo Dual as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter (W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (D) in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where (P) is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Test level
Compliance level Electromagnetic Environment - Guidance
IEC 60601-1-2
3 Veff from 150KHz
3 Veff from 150KHz
to 80MHz
to 80MHz
3 Veff from 80MHz
3 Veff from 80MHz
to 2.5GHz
to 2.5GHz
b
. Interference may occur in the proximity of equipment marked with the
Separation distance according in frequency of transmitter (m)
From 150kHz to 80MHz
d= 1,2P
0,12
0,38
1,2
3,8
12
Portable and mobile RF communication
equipment should be used no closer to any part
of the I-Endo Dual, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
d = 1,2P from 150KHz to 80MHz
d = 1,2P from 80MHz to 800MHz
d = 2,3P from 800MHz to 2.5GHz
where P is the maximum output power rating of
the transmitter in watts (w) according to the
transmitter manufacturer and recommended
separation distance in meters (m).
Dual
From 80MHz to 800MHz
d= 1.2P
0,12
0,38
1,2
3,8
12
a
should be less than the
From 800MHz to 2,5GHz
d= 2,3P
0,23
0,73
2,3
7,3
23
24