- Ask patients and users if they have an implanted device before using this product. Explain
the circumstances to them
- Weigh the risks and benefits and contact your patient's cardiologist or appropriate
qualified healthcare professional prior to performing the treatment
- Keep this product away from implanted devices
- Make appropriate emergency provisions and take immediate action if patients become ill
- Symptom including a raised heartbeat, irregular pulse, and dizziness may signal problems
with a cardiac pacemaker or ICD.
Electrical connection:
- Your device must be connected to the electric power supply by a certified dental installation
technician.
- The electric supply to which the device is connected must comply with the standards in force in your
country.
Using the device:
- Do not use the device if it appears to be damaged or faulty.
- Turn the device off before unplugging the power cord.
- To unplug the power cord, grip the cord plug and hold the wall socket.
- When the device is not to be used for a long period of time, unplug the device from the electric
supply.
- Do not move the device during use.
Environment:
- Do not cover the device.
- Do not immerse the device in liquid, and do not use it outdoors.
- Do not tilt the device at an angle greater than 5°.
- Do not place the device near a heat source.
- Make sure that the cords are not in a traffic path.
- The device should be stored in its original packaging in a safe place.
- The device is not designed for use in the presence of anesthetic gases or any other flammable gas.
- Do not expose the device to water vapor, or splashes.
- The device is not designed to work near ionizing radiation.
- Do not insert metal objects into the device (risk of electric shock, short-circuit or emission of
hazardous substances).
Do not insert anything inside the motor
The "Motor" is supplied Non-Sterile!
Use only power adaptor provided (chapter 3)
Use for intended purposes only. Failure to observe the operating instructions
may result in the patient or user suffering serious injury. Before using this
product, make sure that you have studied and understood the operating
instructions.
Do not install where there is a risk of an explosion. The systems are not
intended for operation in the presence of flammable anesthetics or gases.
Do not disassemble or alter the system motor, console, or footswitch.
Connect mains cord to a properly outlet only.
Never touch files when they are still rotating.
Contra-angle should only be attached when the motor has stopped running.