3 INSTALLATION
3.2.5

Wiring of a compact version with relays, with cable glands

Lift the transparent lid after having unfastened the screw. Remove the cover of the device by
unscrewing the 4 screws, pass the cables through the cable glands and connect according
to the pin assignment below.
Always seal the unused cable gland using the supplied obstructor to
ensure the tightness of the device.
Unscrew the cable gland nut, insert the obstructor and screw the nut back
onto the cable gland.
L+ (12-30 VDC)
4-20 mA output
Strap to be removed if the
4-20 mA output is used
Fit here the cables
using the supplied
cable clips
(see § 3.2.3)
1 2 3 4 PE
Connection relay 2
(see ex. fig. 3.9)
Fig. 3.8
Wiring of the 8206, compact version, with relays, with cable glands
Connecting the 4-20 mA output of transmitter 8206 to a PLC and connection example for
the relays, see fig. 3.9.
1 2 3 4
NC
L+
L-
PE
Without
With
Iout
L+
L-
PE
Supply
12..30Vdc
3A/230VAC
PT SENSOR
Relay cables must be fitted here to the board using
the supplied cable clips (see § 3.2.3)
E-15-
REDOX TRANSMITTER 8206
Earth (internally connected to the earth plug).
Connect the shielding of the power supply cable
Sourcing / sinking mode switch
CURRENT
SOURCE
SINK
PE
redox sensor connection
PH SENSOR
Temperature sensor connection
Connection relay 1 (see ex. fig. 3.9)