Establishing a wireless connection
4.3

Positioning the device

CAUTION
!
Before positioning the devices, a bump test (see
chapter 3.4 on page 40) must be conducted on every
device.
The device may heat up to a temperature outside the
specifications if it is used in very hot conditions and
under direct sunlight at the same time. This may result
in device failure or a reduced service life for some
device components. Dräger recommends placing the
device in the shade under these operating conditions.
NOTICE
i
i
The gas is supplied at an angle of 360°. If required, use
base (order no. 83 20 645) to raise the detection
position by approx. 30 cm.
When positioning the devices, ensure that sub-
networks are avoided.
Use with constant vibration is not recommended.
1. Switch on the first X-zone 5000 (see chapter 3.1 on
page 38) and position at the relevant location.
2. Switch on the second X-
zone 5000 and position at
a distance so that the
communication LED is lit
in green.
The wireless connection
is established and the
communication
LED
indicates the connection
to at least one device
within radio range.
WARNING
!
If the communication LED is lit in red, the distance from
the next device must be reduced.
Position other X-zone 5000 devices using the same
process.
NOTICE
i
i
Dräger recommends performing a connection test
(see chapter 3.4.3 on page 41) after positioning all the
devices.
48
Examples of use:
Scenario A:
Monitoring of industrial tanks (wireless alarm chain)
A
Scenario B:
Area monitoring with pump
B
01333112.eps
Scenario C:
Radio monitoring of pipelines (wireless alarm chain)
C
01833112.eps
01933112.eps
02033112.eps
Dräger X-zone 5000