Télécharger Imprimer la page

Dräger X-zone 5000 Notice D'utilisation page 58

Masquer les pouces Voir aussi pour X-zone 5000:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 67
Faults, cause, remedy
Fault
Horn is too quiet.
Optical alarm signals not visible or poorly
visible.
Life signal is not functioning.
Gas detection instrument not detected.
X-am 5000/5600 switches off quickly; not
powered.
Device defect displayed.
58
Cause
Volume is set too low.
Configuration or pattern incorrectly set.
Configuration is set incorrectly.
IR interface dirty.
Incompatible gas detection instrument.
Incorrect software version in the gas
detection instrument.
IR interface faulty, clip on gas detection
instrument not correctly positioned.
Holder is not correctly positioned.
Power contacts dirty or damp.
X-am 5000/5600 removed during
operation from X-zone 5000.
Poor contacting with X-am 5000/5600.
Remedy
Set the volume using the Dräger CC-
Vision PC software (see chapter 10 on
page 56).
Configure the optical alarm signals using
the Dräger CC-Vision PC software (see
chapter 10 on page 56).
Configure the life signal using the Dräger
CC-Vision PC software (see chapter
10 on page 56).
Clean IR interface.
Use X-am 5000/5600.
Have the software update performed by
DrägerService.
Close the clip on the X-am 5000/5600.
Re-position holder and check that it is
correctly fitted.
Clean power contacts.
Acknowledge alarm on X-zone 5000,
switch off X-zone 5000.
Clean power contacts on the X-zone and
X-am 5000/5600.
Dräger X-zone 5000

Publicité

loading