Use
3.1.1

Charging the Rechargeable Batteries

WARNING
!
Explosion hazard! To reduce the risk of flammable or
explosive atmospheres igniting, it is essential that the
warning notices below are observed:
Do not charge underground or in explosion-hazard areas!
The chargers are not designed in accordance with the
regulations for firedamp and explosion protection.
Charge power packs type HBT 0000 or HBT 0100 with the
appropriate Dräger charger. Charge single NiMH cells for
battery
holder
ABT
manufacturer's specifications. Ambient temperature during
the charging procedure: 0 to +40 °C.
NOTICE
i
i
Even if the instrument is not in use, Dräger recommends that
you store it in the charger (charger module X-am 1/2/5000,
Order No. 83 18 639).
18
All manuals and user guides at all-guides.com
0100
in
accordance
with
Insert the switched off instrument into the charger.
Display LED on the charger:
To maintain the lifetime of the batteries, charging is only performed
within a temperature range of 5 to 35 °C. Outside this temperature
range, the charging is automatically interrupted and resumes
automatically after the temperature is within the range again.
The charging time is typically 4 hours.
A new NiMH power pack reaches its full capacity after three
the
complete charge/discharge cycles.
Never store the instrument for extended periods without being
connected to a power source (maximum of 2 months) because the
internal buffer battery will drain.