Intended use

The Dräger Prestor 5000 is designed as a table-top device for
use in respiratory protection workshops, maintenance centres
or test labs.
The test device can be used to conduct the following tests:
Negative pressure leak test
Opening pressure of the exhalation valve
2.4
Symbol explanation and type
identification
Strictly follow the instructions for use
Medium pressure input for 4-10 bar
USB connection
Mains connection (24 Volt)
24V DC
3

Commissioning the test device

3.1

Instructions on using the test device

To ensure safe and accident-free work with the test device,
observe the following points when working with the test device:
The test device must only be connected to the power
supply using the power pack supplied.
Before commissioning, the test device must be attached to
the workstation using the table mount.
The test device must be protected against ingress of water
and moisture. Ensure that no moisture from splash water or
leaking water can ingress into the test device.
The housing is only permitted to be opened by trained
personnel.
The test device must be sent to Dräger for inspection once
per year. Keep the packaging supplied for future shipping.
Dräger Prestor 5000
3.2

Preconditions for use

The use of the test software requires proficiency in MS Windows.
The PC on which the test software is to be installed must meet
the following minimum requirements:
Processor: 1.6 GHz or higher
RAM: 2 GB or more
Free disk space: >500 MB
Operating system: Windows 7 or Windows 8
1 free USB port
The monitor resolution should be set to 1024 x 768.
3.3

Prior to first use

Before the test device can be put into operation, the following
steps must be carried out:
1. Install the test software on the PC and create the data
(see chap. 3.3.1 on page 11).
2. Set up the test device (see chap. 3.3.2 on page 12).
3.3.1
Installing the test software on the PC and creating
the data
NOTICE
i
i
For installation, the user must have administrator
rights for the PC.
1. Start "Protector.exe" from the USB memory stick.
2. Follow the instructions in the setup program.
The required program components will be installed in the
selected folder and an entry created in the start menu.
3. Observe the information displayed during installation.
4. Re-start the PC and log in as user "Chief".
5. Import the required device data with the test settings from
the USB stick. The following data sets are available:
For the EU database in the directory EU:
mask_EU.zip (device data for full face masks)
6. In the test software, create additional users as well as
respiratory protection products, customers and locations
as required (see "First steps" in the help pages).
NOTICE
i
i
On delivery, the test software includes the following
user:
"Chief" with password "Createdata"
Information on the test software is also provided on the
help pages.