7.5 SETUP Menu (determining basic settings)

Description
Under this menu the user can carry out or alter different basic settings himself.
Operation
SETUP-Taste
Autostart
Abort at Q = 0
Pump pre-run time
Date / time
Bus address
Operating hours
Battery voltage
BS pressure supply
Additional test
Autostart:
Abort at Q = 0
HYDAC Filtertechnik GmbH
yes
The FCU starts automatically after switching on with
measuring in the pre-set measurement mode (Autostart
function)
no
Start of measurement must be triggered manually.
yes
If the flow rate through the sensor falls to the value "0",
then measurement is aborted and the FCU switched off.
no
If the flow rate through the sensor falls to the value "0",
then measurement is aborted, but the FCU still remains
switched on and continues measuring when sufficient flow
is available.
E
Doc.: 3109776 Ed.: 01/13.02.2002