9 Maintaining your STATIM

9.7 Maintaining Fluid Levels

1. The water reservoir level is continually monitored by your STATIM. If the reservoir is low, a red X will
appear on the water icon
confirm that it is a water level issue and not a water quality issue.
2. If the reservoir is low, a
3. To fill the reservoir, use steam-processed distilled water or SciCan approved water sources containing
less than 5 ppm total dissolved solids (having conductivity of less than 10 μS / cm). Remove the cap
from the top of the unit and fill the reservoir. We recommend using a funnel to minimize spills. Each
time you refill the reservoir, empty the waste bottle and refill with water to the MIN line. Empty the
waste bottle often to avoid unpleasant odors and discoloration of the contents. (A low-level chlorine-
free disinfectant, prepared according to the manufacturer's instructions, may be added to the waste
bottle to remedy this situation).

9.8 Reading Water Quality

1. The water quality is continually monitored by your STATIM to ensure the optimal quality of water is
being used. If the water quality is beyond the tolerances specified above, a red X will appear on the
water icon
of the cycle select screen. Press the icon to go to the next screen to confirm that it is a
water quality issue and not a water level issue.
2. If the water quality not suitable,
3. Using the drain tube (see Section Priming the Pump, Figure 6) empty the contents of the reservoir into
the water container and replace with steam-process distilled water containing less than 5 ppm total
dissolved solids (having conductivity of less than 10 μS / cm).

9.9 Using On-Screen Instructions

The STATIM G4 on-screen help instruction can be viewed by following these steps:
1.
2. Scroll to
Instructions
3. Choose the Instructions you would like to view and select.
of the cycle select screen. Press the icon to go to the next screen to
will appear next to WATER LEVEL.
will appear next to micro S. and parts per million values.
and select.