3 Setting up your STATIM

3.4 Filling the Water Reservoir

When filling the reservoir, ensure you only use steam-processed distilled water or SciCan approved water
sources containing less than 5 ppm total dissolved solids (having conductivity of less than 10 µS / cm).
The impurities and additives in other water sources will cause an error reading on the LCD.
To fill the reservoir, follow these steps (see Figure 5):
1. Remove the reservoir cap
2. Pour steam-process distilled water into the
reservoir until almost full (a maximum of 4L /
1 U.S. gal). Use a funnel to avoid spillage.
3. Replace and secure the cap.

3.5 Priming the Pump

To prime the STATIM pump, follow these steps
(see Figure 6):
1. Move the unit to the edge of the work surface.
The front leveler feet should be approximately
12 mm/0.5˝ from the edge.
2. Lift the front left corner of the unit upward and
remove the drain tube
the underside of the unit.
3. Pull the drain tube outward so the free end can be
positioned over a water container.
4. Fill the reservoir with steam-process distilled
water.
5. Remove the plug
tube and allow water to drain from the tube into a container for 30 seconds. When the water flows in a
steady stream, replace the stopper.
6. Lift the front left corner of the unit upward and reinsert the tube into the clip on the underside of the
unit. Push the excess length of tubing back into the space provided.
Make sure the plug in the drain tube is secured.
After installation, and before sterilizing any instruments, run two Wrapped cycles. For further
instructions, see Section 5.1 Preparing the Unit for Use.

3.6 Shipping the Unit

Before you move the unit, you will need to drain the reservoir. To do so, follow these steps:
1. Place a water container below the unit.
2. Using the drain tube (see Section 3.5 Priming the Pump, Figure 6) empty the contents of the reservoir
into the water container.
3. Remove any remaining water from the reservoir with a non-linting, absorbent towel.
4. Screw-in the three leveler feet found underneath the unit.
5. Repack the unit in the original packing materials and include all accessories originally included with
the unit.
6. Specify heated and insured shipping.