Caster Forks; Footrest; Wing Extensions - Sunrise Medical Quickie All Court Manuel D'instruction

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 41
E N G L I S H
I. CASTER FORKS
Caster Wheel Replacement
a. Loosen the bolt and remove the caster wheel and the two washers.
b. Position new caster wheel into fork.
c. Carefully position washer on each side of caster wheel.
d. Insert bolt. Tighten securely.
J. FOOTREST
NOTE: Use a torque setting of 100 in./lbs. when adjusting footplate.
1. Height Adjustment
a. Remove footrest bolt and slide footrest up or down inside tube to correct position.
b. Tighten bolt securely.
2. Footplate Angle Adjustment
a. Loosen flathead screws.
b. Reposition footplate to desired angle and retighten screws.
K. WING EXTENSIONS
Wing Adjustments
a. Loosen bolts on wing clamps.
b. Slide wing out or in to desired location.
c. Retighten bolts into clamps.
L. ANTI-TIPS
Sunrise recommends anti-tip tubes for all wheelchairs.
Caster Height Adjustment for the Single and Double Integral Anti-tip
The anti-tip height can be adjusted by repositioning the spacers within the caster journal.
Each caster journal is equipped with four 1/8" thick spacers (1).
a. To lower the casters remove desired spacers and position them on the bottom
of caster journal (2).
b. To raise the height of the caster remove desired spacers and move to the top of the caster
journal. With the double integral make sure both casters are at the same height.
c. Additional height may be achieved using the second mounting hole in fork.
930493 Rev. A
Single Integral
Double Integral
(Optional)
16