Fig. 11:
D184B112U04
FXL5000 (Miniflow)

Connection diagram, examples of connections for peripherals

Variant 01 (without display) or 03 (with display)
(pulse output passive, optooupler or
contact output passive, optocoupler)
internal
external
Pin 3
Pin 4
Pin 2
Not assigned
Pin 1
Not assigned
Variant 05 (with display)
(pulse output passive, optocoupler and contact input passive, optocoupler or
contact output passive, optocoupler and contact input passive, optocoupler)
internal
external
R
Pin 3
Pin 4
Pin 2
R
I
Pin 1
R
= 2 kΩ
I
Auxiliary power
(24 V AC/DC)
Pin 1
24 V AC/DC
Pin 2
GND
Pin 4
Functional earthing
Technical data of signal outputs
Pulse output, passive
(optocoupler data:)
16 V ≤ U
≤ 30 V; 0 V ≤ U
≤ 2 V;
CEH
CEL
0 mA ≤ I
≤ 0.2 mA ; 2 mA ≤ I
CEH
CEL
fmax = 20 pulse/sec; pulse width 20 ms – 2550 ms
* Contact output, passive, optocoupler
The function of the optocoupler for the pulse output
can be changed over using software.
It may either function as a pulse output or contact output.
The changeover function is only available with the version with a
display!
One of the following statuses can be signalled in the "contact out-
put" function:
Inlet/outlet signalling, min./max. contact, system alarm.
If working with the version without a display, the opto-electronic
coupler function can only be set as "pulse output" or "inlet/
outlet signalling".
Examples of connections for peripherals, signal output data
R
*
B
Pulse output/
contact output*
U
CE
*R
---------- -
B
I
CE
*
B
Pulse output/
contact output*
24 V+
0 V
Current output
can be set to between 0/4 and 20 mA; Load ≤ 600 Ohm
Contact input
≤ 220 mA
The input is considered as actuated at an external
voltage of between 16 V ≤ U
It is considered as non-actuated at a voltage level of between
0 V ≤ U
contact input is: Ri = 2 kOhm.
Auxiliary power
16.8 V AC - 26.4 V AC and/or 16.8 V DC - 31.2 V DC
Ripple: 5 %
Power consumption< 5 W
5 Electrical connection
Variant 02 (without display) or 04 (with display)
(pulse output passive, optocoupler and
current output 0/4 −20 mA or contact output passive,
optocoupler and current output 0/4−20 mA)
internal
external
R
B
Pin 3
Pin 4
+
Pin 2
0/4 - 20
Pin 1
≤ 30 V DC.
CEH
≤ 2 V DC. The internal resistance of the
CEL
*
Pulse output/
contact output*
GB 15