K. REDOX CALIBRATION PROBE
1
5
Press
until the
message "CALIBRATION"
is displayed
The message
"Press_CAL"
"Buffer sol 465mV"
is displayed
9
Combipool-pHRedox_(Multi) - DOC015452 - Ind. A03 - 23/12/2020
Marque, Modèles et Brevets déposés - Produits exclusifs
2
6
Press
The message
"Wait 60 s"
"Buffer sol 465mV"
is displayed
10
Press
to exit and
save the data
3
Wash the probe
7
If the message
« 465mV Quality' 100% »
is displayed, go to step
8.
If the message
"Change the probe"
is displayed, press
to exit.
Change the probe and
return to step 1.
11
Displays the redox
measure in the pool
4
Hold the probe in the buffer
solution
8
Wash the probe