J. PH PROBE CALIBRATION
Before calibration, check that the buffer solutions have not expired. The probe should be calibrated at least
yearly.
1
5
Press
until the
message "CALIBRATION"
is displayed
The message
"Press_CAL"
"Buffer sol pH 7"
is displayed
9
Hold the probe in the buffer
solution
13
Combipool-pHRedox_(Multi) - DOC015452 - Ind. A03 - 23/12/2020
Marque, Modèles et Brevets déposés - Produits exclusifs
2
6
Press
The message
"Wait 60 s"
"Buffer sol pH 7"
is displayed
10
Press
The message
"Wait 60 s"
"Buffer sol pH 4"
is displayed
14
Press
to exit and
save the data
3
Wash the probe
7
If the message
« 7pH Quality' 100% » is
displayed, go to step 8.
If the message
"Change the probe"
is displayed, press
to exit.
Change the probe and
return to step 1.
11
If the message
« 4pH Quality' 100% » is
displayed, go to step 12.
If the message
"Change the probe"
is displayed, press
to exit.
Change the probe and
return to step 1.
15
Displays the pH in the pool
4
Hold the probe in the buffer
solution
8
Wash the probe
12
Wash the probe
Note :
When the EASY CAL
function is selected,
calibration is only
performed for 1 point at
pH7.
Copyright BIO-UV
Page 37