4.1.2 Description of functions

Menu
Parameter
Set parameter: Setting of the measuring parameter
(CAL = rEF.S)
46
Values
Meaning
Setting of the cell constant: Cell constant range
0.01
e.g. ultrapure water,
electrodes with K ~ 0.01
0.1
e.g. ultrapure water,
electrodes with K ~ 0.1
1
e.g. electrodes with K ≈ 1, K ≈ 0.55,
K ≈ 0.42
10
e.g. electrodes with K ≈ 10
Setting of the cell constant: Multiplication
factor
0.3800 ...
Multiplication factor of the cell
1.500
constant
Cell constant K = CELL rAnG * CELL FACt
Setting: Display range / resolution
Auto
Automatic resolution of the
measuring range
Manual
Manual selection of the resolution
for the measuring range
Cell constant comparison using reference
solutions
Edit
Manual adjustment to the reference
value
REF.S
Selection from standard reference
solutions
Selection from standard reference solutions for
automatic adjustment of the cell constants
147 µS/cm
Reference solution (c = 0.001 M KCl)
1,413 µS/cm
Reference solution (c = 0.01 M KCl)
2,760 µS/cm
Reference solution (c = 0.02 M KCl)
12.88 mS/cm Reference solution (c = 0.1 M KCl)
50 mS/cm
Seawater comparison solution
111.8 mS/cm Reference solution (c = 1 M KCl)
*
*
*
*
SD_325_Con_1 06/2020