Technical operating data
Radio approval:
13.1
Technical data when operated with CO
6800482
Operating pressure
700 hPa to 1300 hPa
range:
Measuring range:
0 to 10000 ppm
Accuracy at 25 °C
±30 ppm, +3 % of measured value
Alarm threshold A1:
1000 ppm
Alarm threshold A2:
2000 ppm
13.2
Technical data when operated with NH
6813735
Operating pressure
700 hPa to 1300 hPa
range:
Measuring range:
0 to 100 ppm
Resolution:
0.5 ppm
Accuracy:
1.5 ppm or ± 10 % of the measured value
(larger value applies)
Response time:
T50 < 30 seconds
Long-term drift:
< ± 10 % of measured value/year
18
EN 62311:2008
EN 300 220-1 V3.1.1
EN 300 220-2 V3.1.1
EN300 328 V2.2.2: (2019-07)
47 CFR Part 15.247
AS/NZS 4268:2017
sensor
2
AL sensor
3
13.3
Technical data ambient sensors
Ambient pressure:
Temperature:
Humidity:
Instructions for use
Accuracy
700 hPa to 1100 hPa = ± 1.0 hPa,
1100 hPa to 1300 hPa = ± 1.5 hPa
Accuracy
0 °C to 50 °C = ± 2.0 °C
-20 °C to 0 °C = ± 2.5 °C
5 % to 95 % RH: Accuracy ±
10 % RH
|