Technical data

Measurement principle
Measuring range
Breath alcohol concen-
tration
Blood alcohol concen-
tration
A conversion factor of 2100 is used to calculate the blood alcohol
concentration using the breath alcohol concentration data (factory
setting, country-specific conversion factors can be configured).
Measuring units
Mg/L, μg/100 mL, μg%, μg/L, g/210 L, g/L, ‰, %, mg/100 mL,
mg%, mg/mL
Ambient conditions
During operation
In storage
Data storage
Sampling
Minimum volume of
breath
Breath flow
Dräger Alcotest
Electrochemical sensor
Up to 30,000 tests (depending on the
configuration)
Warning that data storage is almost full
at 90 %
Default: FIFO (oldest test is overwrit-
ten)
Optional: Device locks at 100 %
®
7000
0.00 to 3.00 mg/L (mass of alcohol
per minimum volume of breath at
34°C and 1013 hPa)
0.00 to 6.00‰ (mass of alcohol per
blood volume or mass of blood at
20°C and 1013 hPa)
–10 to +55 °C
10 to 100 % r.h. (non-condensing)
600 to 1300 hPa
–20 to +60 °C
15 to 75 % r.h. (10 to 100 % r.h. <48
h)
>1.2 L (default setting)
>10 L/min (default setting)
Technical data
27
|