Technical Data

Extract: For details refer to the Technical Handbook.
General Data
ambient conditions:
IP rating
power consumption
warm-up time (after power-up)
electrical connection
CE marking
dimensions
weight
approvals:
1) The year of construction is indicated by the third letter of the serial number on the type plate: Y = 2007, Z = 2008, A = 2009, B = 2010, C = 2011, etc.
Example: serial number ARYH-0054, the third letter is Y, so the year of construction is 2007.
20
o
for operation
–40 to +77
C (–40 to +170
o
during storage
–40 to +85
C (–40 to +180
IP 66 and IP 67, Nema 4X
5.6 W (typical), <7 W (maximum)
1 minute
cable diameter 7 to 12 mm, wire cross-section max. 2.5 mm
Devices and protective systems for use for the intended purpose in explosion-hazard area
(Directive 94/9/EC)
Approx. 160 mm x Ø89 mm
Approx. 2.2 kg (without accessories)
ATEX
Type: IDS 01xx
Dräger Safety • 23560 Lübeck • Germany
0158
II 2G Ex d IIC T6/T4 – DEMKO 07 ATEX 0654417X
II 2D Ex tD A21 IP65 T80
o
C d Ta d +40
–40
measuring function for the explosion protection (4-20 mA signal output only) - BVS 08 ATEX G 001 X
Type 334: 0 to 100 % LEL methane, 0 to 100 % LEL propane, 0 to 100 % LEL ethylene, 0 to 100 % vol. methane
Type 340: 0 to 100 % LEL propane, 0 to 5000 ppm propane, 0 to 100 % LEL methane
Year of construction (via serial number)
IECEx
Type: IDS 01xx
Dräger Safety • 23560 Lübeck • Germany
Ex d IIC T6/T4 – IECEx UL 07.0009X
Ex tD A21 IP65 T80
o
C d Ta d +40
–40
Year of construction (via serial number)
UL (Classified)
Class I, Div. 1, Groups A, B, C, D / Class II, Div. 1, Groups E, F, G / Class I, Zone 1, Group IIC
T-Code T6/T4, –40
9 to 30 V DC, 9 W - Type 4x
CSA (C-US)
Class I, Div. 1, Groups B, C, D / Class II, Div. 1, Groups E, F, G
T-Code T6/T4, –40
C22.2 No. 152
9 to 30 V DC, 9 W - Type 4x
o
F), 700 to 1300 mbar, 0 to 100% RH
o
F), 700 to 1300 mbar, 0 to 100% RH, non-condensing
o
o
C/T130
C – PTB 07 ATEX 1016
o
o
C/+80
C
1)
o
o
C/T130
C – IECEx PTB 07.0016
o
o
C/+80
C
1)
o
C d Ta d +40
o
o
C/+80
C
o
C d Ta d +40
o
o
C/+80
C
2
2
or 2 x 1 mm