Télécharger Imprimer la page

Dräger X-node Notice D'utilisation page 15

Publicité

Les langues disponibles
  • FR

Les langues disponibles

8
Getting started
8.1
Mounting Dräger X-node
CAUTION
The sensing head can be damaged.
To ensure product safety even when an external power supply is
connected to the micro USB socket, only an approved, commercially
available USB power supply unit may be used that does not exceed
the parameters 5.25 V / 3 A / 15 W.
► The sensing head may only be used with USB power supply units
that comply with the USB 3.1 (Gen2) specification or lower.
CAUTION
The sensing head can be damaged.
No IP rating applies during the charging process.
► The sensing head must be protected against spray water and dust
while the rechargeable battery is charging.
The internal rechargeable battery of the sensing head must be
charged as follows prior to the installation of the sensor in the sensing
head (3):
1. Remove the seal for the micro USB port (6).
2. Connect the plug of a micro USB cable to the micro USB port (6)
of the sensing head.
3. Connect the micro USB cable to a compatible power supply unit.
4. Connect the power supply unit to the mains power supply.
5. When the charging process is complete, the micro USB port must
be resealed with the seal.
For optimum measurements the sensing head must be installed at the
centre of the area to be monitored. The sensing head must be
installed so that it is accessible for maintenance work and suspended
from the mounting lug (1) or installed upright on the bayonet ring (5).
Instructions for use
|
Dräger X-node
The sensing head must be mounted vertically with the sensor (5)
pointing down, ±30 ° at a location with low vibration.
8.2
LoRa gateway connection
For successful data exchange between sensing head and gateway,
the gateway must be within range. Additional information on LoRa
can be found in the device's technical manual.
|
en
Getting started
15

Publicité

loading