3.4.2
Checking the compressed oxygen cylinder
If the compressed oxygen cylinder has come into contact with
corrosive media (e.g. acid), it has to be depressurised and Dräger
must be consulted to clarify the subsequent course of action.
If the compressed oxygen cylinder had to withstand a heavy
impact(for example, because it was dropped), then it must be
depressurised and sent to Dräger for assessment.
Compressed oxygen cylinders must be discarded in the following
cases:
The
compressed
deformations such as dents in the material, grooves or pulping
that are longer than 0.78" (20 mm) or deeper than 0.009"
(0.25 mm).
The compressed oxygen cylinder shows signs of the
paintwork or composite material flaking or blistering or other
structural damage.
The compressed oxygen cylinder shows signs of abrasive
wear.
The cylinder material, label or paint are burnt or blackened or
the paint has melted.
A clear allocation of the compressed oxygen cylinder is not
possible because the label is illegible.
If a clear-cut assessment of the damage is not possible, then
Dräger should be consulted to clarify the subsequent course of
action.
If the date for the periodic test has been exceeded, the
compressed oxygen cylinder may not be filled or used. A periodic
test must be arranged.
Compressed Oxygen Cylinders
oxygen
cylinder
has
permanent
4
Maintenance
4.1

Maintenance intervals

Work to do
Check filling pressure
Fill the compressed oxygen cylinder
2)
Periodic test
4)
Major overhaul of the valve
1)
see Section 3.2.1 on Page 5
2)
by an accredited inspection authority
3)
observe national guidelines
4)
by Dräger or an authorized service technician trained by Dräger
Maintenance
X
X
1)
X
3)
X
X
7