Recommended Inspection And Maintenance Intervals; Cleaning And Disinfecting - Dräger X-plore 6000 Notice D'utilisation

Masquer les pouces Voir aussi pour X-plore 6000:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 22
4
Maintenance
4.1
Recommended inspection and maintenance
intervals
The following information is taken from BGR 190, one of the
Regulations of the Employer's Liability Insurance Association in
Germany. Observe national directives.
Type of work to be
performed
Check by equipment wearer

Cleaning and disinfecting

Visual inspection, function
and leak test
Replace exhalation valve
disk
Replace speech diaphragm
Replace O-ring of the speech
diaphragm
1)
for airtight packed full face masks, otherwise every 6 months
2)
for airtight packed full face masks every 2 years
Dräger X-plore 6000
X
1)
X
X
2)
X
X
X
X
(approved by Dekra Exam)
4.2
Cleaning and disinfecting
Clean and disinfect the full face mask after every use.
WARNING
!
The mask body must be cleaned sufficiently thoroughly that no
residues from skin care products remain on it.
Otherwise the elasticity in the sealing area can be lost and the
mask will no longer sit correctly.
CAUTION
!
Do not use any solvents (e.g. acetone, alcohol) or cleaning
agents containing abrasive particles for cleaning and
disinfection. Only use the prescribed process and use the
cleaning and disinfection agents given. Other agents, dosages
and contact times may damage the product.
The undiluted agents are damaging to health if they come into
direct contact with the eyes or skin. For this reason, wear
safety goggles and protective gloves when working with these
agents.
1. Remove any accessories and clean them separately.
2. Clean all parts with lukewarm water with the addition of Sekusept
X
Cleaner using a soft cloth (temperature: max. 30 °C, concentration
depending on the degree of contamination: 0.5 - 1 %)
X
3. Rinse all parts thoroughly under running water.
4. Prepare a disinfection bath consisting of water and Incidin
(temperature: max. 30 °C, concentration: 1.5 %)
5. Place all parts that require disinfection into the disinfection bath
(duration: 15 minutes).
6. Rinse all parts thoroughly under running water.
7. Allow all parts to dry in the open air or in a drying closet
(temperature: max. 60 °C). Do not expose to direct sunlight.
®
1)
Sekusept
is a registered trademark of Ecolab Deutschland GmbH
®
2)
Incidin
is a registered trademark of Ecolab USA Inc.
Maintenance
®
1)
.
®
Rapid
2)
.
17

Publicité

Table des Matières
loading

Table des Matières